The OpTech®- O₂ Platinum measures oxygen permeation, headspace, dissolved oxygen, and package leak. It is the ideal multi-application oxygen analyzer for food, beverage, pharmaceutical and medical application where measuring oxygen and understanding its effect on product and product shelf life is critical.

 
  • Ergonomic
  • No bulky fiber optic cables
  • Lightweight and portable
  • Reusable, easy to place sensors – no glues needed
  • Non-consuming sensor - designed for limited headspace
  • Visible, non-UV light source for simple, targeted reading
  • ImPULSE™ sensor for opaque packaging materials
  • Packaging oriented software
  • No gas calibration needed for the reader
  • Accurate readings through colored packaging material
  • One calibration for all adhesive sensors
  • Non-consuming sensors make it ideal for long term O₂ studies

Detector and Base

  • Warm-up time: 20 minutes
  • Detector dimensions: Width: 1.3 in 3.30 cm Height: 1.9 in 4.83 cm
  • Depth: 9 in 22.86 cm (with needle), 6 in 15.24 cm (without needle)
  • Base Dimensions: Width: 4.8 in 12.19 cm Height: 2.7 in 6.8 cm
  • Depth: 10 in 25.40 cm
  • Measurement method: Epifluorence Confocal
  • Power: Standard Power USB Port (2.5 watt)
  • Operating temperature: 10 – 35C
  • Operating humidity: 0 – 100% non-condensing
  • Compliance: CE/CSA/UL
  • PDF report options: through program from computer

Sensors (Adhesive and ImPULSE™)

  • Application:
  • Adhesive: Sensor is inside package
  • ImPLUSE Sensor is external
  • Repeatability Adhesive (Certified): ±0.015% (150 ppm) O₂ or 3% of reading, whichever is greater
  • Repeatability ImPULSE (Certified): ±0.05% (500 ppm) O₂ or 3% of reading, whichever is greater
  • Range Adhesive:
    • 0.001% (10 ppm) to 25% O₂ Permeation Mode
    • 0.015% (150 ppm) to 25% Headspace Mode
  • Range ImPULSE: Range ImPULSE 0.05% O₂ (500 ppm) to 25% O₂
  • Dissolved Oxygen Range: 0.006mg/L to 10.5mg/L
  • Warm up time: None
  • Adhesion: Sensors come ready to apply
  • Operating temperature: 5 – 40C
  • Operating humidity: 0 – 100% Sensors are designed to be immersed

Needle Sensor

  • Application: Sensor in needle must be 100% in volume
  • Repeatability (Certified): ±0.015% (150 ppm) O₂ or 2% of reading, whichever is greater
  • Range: 0.015% (150 ppm) to 25% Headspace mode
  • Warm up time: None
  • Operating temperature: 5 – 40C
  • Operating humidity: 0 – 100% non-condensing. Needle must not get wet

Computer Tablet and Case

  • Dimensions: (w x d x h) 5.9 x 9.21 x 0.61 in (15 x 23.4 x 1.5 cm)
  • Weight: Starting at 1.5 lb (0.69 kg)
  • Operating system: Windows® Professional 32
  • Ports: 1 USB 2.0
  • Carrying Bag: 9 x 2.5 x 12 in (22.86 x 6.35 x 30.48 cm )

CalCard®

  • Accuracy: ±2% or ±150 ppm, whichever is greater

CalVial®

  • Accuracy: ±150 ppm