The OX-TRAN Model 702 is ideal for conducting shelf life studies, quality control, and other research and development applications. It is a 7 cell instrument and tests both flat films and finished packages. Only MOCON instruments comply with ASTM D3985.

  • Complies with standards:
    • ASTM D3985
    • ASTM F1307
    • DIN 53380-3 films and JIS K-7126 films
    • ISO CD 15105-2
  • Seven test cells
  • Quick start Function
  • Patented Coulox® sensor
  • Pneumatic clamping push button test cells
  • Removable cells
  • Operator friendly software
  • Automatic electronic flow control
  • Test up to seven films or packages simultaneously
  • No calibration required
  • Film and package testing in any combination
  • Easy sample preparation
  • OTR Test Range: (see Note 1)
  • Test Temperature Range: 5C – 50C with computer controlled bath
  • Relative Humidity Conditions: Dry
  • Sensor: Red COULOX Sensor
  • Test Sample Size:
    • Films: 4.25 in x 4.25 in (10.8 cm x 10.8 cm)
    • Packages: Up to 3 liters per package
  • Number of Test Cells: Seven-50 cm2
  • Expandable up to 10 modules: (70 test cells)
  • Automatic Temperature Control: (Standard)
  • Computer, Monitor, Printer and WinPerm® Permeability Software (Standard)
  • Operating System: Windows® 2000
  • Automatic Digital Barometric Pressure Compensator (optional)
Note #1
Sample cc/(m2 · day) cc/(100 in2 · day) cc/(pkg · day)
Unmasked 0.01 – 200 0.00025 – 1.0 0.00005 – 1.0
Masked 0.1 – 2000 0.007 – 129