Oxygen Permeation Analyzer for Testing Films & Finished Packages

The OX-TRAN Model 2/40 Oxygen Permeation Analyzer is specifically designed to measure the oxygen transmission rate (OTR) of whole packages. Your operators will now spend less time setting up and monitoring package permeation tests, allowing them to achieve more in less time.

The new controlled chamber instrument has 4 independent test cells allowing multiple package tests to be run at the same time. Various package test adaptors are available to allow easy mounting of trays, bottles, flexible pouches, corks, cups, caps and more. The temperature and humidity-controlled chamber guarantee an accurate and reliable test gas condition for every test.

The OX-TRAN 2/40 series of oxygen permeation analyzers are offered in three models:

  • Model 10X
  • Model L
  • Model H

Contact us to learn more or to set up a live demo!

Industry Standards:
  • ASTM F1307 packages
  • ASTM D3985 films
  • JIS 7126-B films
  • ASTM F1927 films (with RH)
  • DIN 53380-3 films
  • ISO 15105-2

Absolute Precision

  • Test entire package in a controlled atmosphere
  • Temperature and humidity-controlled chamber
  • No calibration required

Optimized Productivity

  • Save on labor hours with easy setup and less in-test monitoring
  • Accelerated test options
  • Simultaneously test up to 4 large packages 

User-Friendly

  • Various test cell adaptors for different package types
  • Simplified sample preparation
  • Touch screen interface
Test Standards
Test Temperature Range 10°C to 50°C ± 0.2°C
Chamber RH Testing Ranges 5% to 90% ± 4% ** Details in Technical Data Sheet
Carrier Gas RH Testing Ranges 0% to 90% ± 4% ** Details in Technical Data Sheet
Technical Specifications
Test Ranges Resolution Repeatability Min Exam Time
cc/(pkg • day) cc/(m2 • day) cc/(pkg • day) cc/(pkg • day) Minutes
Model 10X
Atmospheric Tested (20.0% O₂) 0.0000025 to 1 0.0005 to 200 0.000001 ± 0.0000025 60
Software compensated to 100% O₂ 0.0000125 to 5 0.0025 to 1,000 N/A
Model L
Atmospheric Tested (20.0% O₂) 0.000025 to 1 0.005 to 200 0.00001 ± 0.00001 45
Software compensated to 100% O₂ 0.000125 to 5 0.025 to 1,000 N/A
Model H
Atmospheric Tested (20.0% O₂) 0.00025 to 1 0.05 to 200 0.0001 ± 0.0001 30
Software compensated to 100% O₂ 0.00125 to 5 0.25 to 1,000 N/A